Two laboratory reports can show apparently reassuring results without describing comparable measurements. In a fictional paperwork example, Report A names its measurement network and accessible part; Report B gives only the tester model and a generic “leakage current” heading. Neither the shared instrument category nor a smaller displayed result closes those missing fields.
Before comparing values, establish what was measured, through which documented method, and in which state. The dossier below is an original document-review aid for purchasing and supplier-quality teams, not an IEC test procedure or certification checklist. It contains no actual charger readings and supplies no acceptance limit. Electrical testing belongs only in a qualified, controlled laboratory; this article does not authorize a setup or fault experiment.
Name the measured path before comparing the result
GW Instek’s GLC-9000 user manual explains measuring devices, or MDs, as body-simulation networks that vary with the type of test. Its opening current-path classification separates current through a person touching equipment, current through a protective grounding conductor, and current involving an applied part. Those distinctions support recording the actual measurand and network. They do not make the manual a complete safety standard or make its medical examples applicable to a charger.
Do not silently rename a protective-conductor measurement as touch current. Preserve the laboratory’s original term, identify the stated path, and ask for clarification when “leakage” is used without qualification. Likewise, do not derive the measurement state from a phrase such as “floating output.” Such wording needs the laboratory’s method-specific explanation, not an assumed safety conclusion.
An input energization waveform is a different subject. Use the input-inrush report review for that evidence rather than mixing it into this dossier.
A tester name does not identify the selected method
Chroma’s 19032/19032-P technical datasheet lists measurement-device options separately from measurement modes, including normal, reverse and single-fault descriptions. This supports asking which options and state were actually recorded, rather than treating an instrument model as a complete method. The datasheet spans multiple applications and legacy standards; its instrument ranges and medical examples are not charger acceptance limits.
Record the applied standard and edition separately from the laboratory method revision and selected network identity. Ask the laboratory to explain the relationship between those references. A familiar standard name on a cover does not establish which case a particular result represents. Neither instrument document selects the required network for the buyer.

Build a paired dossier with explicit missing fields
Use the following dossier as a comparison template, not as a test setup. The cells describe evidence to transcribe from each report; they are not completed findings. Attach report revisions and page references so another reviewer can reconstruct the comparison without guessing.
| Evidence field | Report A record | Report B record |
|---|---|---|
| Sample identity | Exact model, suffix and sample ID | Corresponding model, suffix and sample ID |
| Method basis | Standard edition and laboratory method | Standard edition and laboratory method |
| Measurand and path | Original measurement term and path | Original measurement term and path |
| Measurement network | Named MD and revision or method reference | Named MD and revision or method reference |
| Accessible part | Laboratory-defined part or point reference | Corresponding part or point reference |
| Detector and output | Declared metric, weighting and filter | Declared metric, weighting and filter |
| Operating condition | Recorded input, load and port state | Recorded input, load and port state |
| Equipment description | Recorded class and earthing description | Recorded class and earthing description |
| Test state | Normal or specified fault-case identifier | Normal or specified fault-case identifier |
| Instrument traceability | Instrument ID and calibration reference | Instrument ID and calibration reference |
| Result interpretation | Results, repeats, uncertainty and cited criterion | Results, repeats, uncertainty and cited criterion |
Give every row a separate comparison flag: same, intentionally different, or missing. Add the consequence, the laboratory question and the person authorized to close it. Keep multiple operating or fault states as separate case records; do not collapse them into one unlabeled maximum. “Not stated” is useful evidence about the report, but not proof that the laboratory omitted the work.
UL Solutions’ Calibration and Variation Requirements, within its own customer program, call for identifiable calibration status and an unambiguous instrument identifier. The document also describes reviewing calibration certificates and data for the measurements the equipment will perform. This is a bounded traceability example, not a universal laboratory rule. Calibration evidence does not establish that a touch-current network, path or state was appropriate.
Separate a method gap from a possible product finding
Classify the consequence before asking for more files. The following documentary cases carry no invented readings and make no finding about a real charger. Their purpose is to preserve the distinction between incomplete comparison evidence and a result needing qualified safety review.
| Documented situation | Comparison disposition | Evidence or owner needed |
|---|---|---|
| Network missing or different | Hold the numerical comparison | Laboratory explanation of the actual MD and applicability |
| Normal compared with fault state | Keep separate cases | Matched-state records or an explanation of the intended comparison |
| Floating-output wording alone | Leave method identity unresolved | Laboratory-defined path, state and method reference |
| Method, path and state match | Compare only equivalent cases | Applicable criterion and qualified interpretation still needed |
| Traceable result exceeds its stated applicable criterion | Escalate a possible product safety finding | Designated safety reviewer and laboratory assessment |
A higher result in a different state is not automatically a failure. Conversely, an adverse result with traceable method evidence should not disappear behind a generic “method unclear” label. Preserve it, identify the cited acceptance basis, and let the responsible reviewer determine the consequence. Purchasing staff should neither invent a replacement limit nor average unlike cases to obtain a reassuring number.
Send a question the laboratory can answer
Reference one unresolved dossier row in each question. Quote the ambiguous term, identify both report revisions, and request the method, network and case references needed to decide whether the results are comparable. Ask how the accessible-part record, instrument identity, uncertainty and stated criterion relate to that case. Request a controlled clarification or amended report, rather than an informal assurance that the charger is “safe.”
Keep the original report, the laboratory’s dated response and the reviewer’s disposition together. A useful closure states which comparison is supported and which cases remain outside it. Resolving paperwork does not itself authenticate the product, establish certification or approve shipment.
For a ChargeKeku sourcing enquiry, request model-bound laboratory report references with the exact configuration and unresolved fields. This is a documentation request, not a promise of testing, certification or a safety pass.
Keep the review on the document side
Do not open a charger, probe mains circuits, interrupt protective earth or create a fault to fill an empty cell. Qualified laboratory personnel must determine any necessary work within their controlled safety procedures. No complete IEC measurement or charger safety standard was obtained for this article, and no network circuit, component value or current threshold is prescribed here.
A later design substitution raises a different question. Transfer it to the owner of configuration changes and requalification decisions with the dossier references attached. This comparison aid does not select retests or authorize omitting them.
Sources
- GW Instek — GLC-9000 User Manual — selected measuring-device discussion and opening current-path classification; not a complete safety standard.
- Chroma ATE — 19032/19032-P technical datasheet — measurement modes and device options, not charger limits.
- UL Solutions — Calibration and Variation Requirements — selected instrument-identity and calibration-scope paragraphs within UL’s customer program.